Applications:
- With liquid filled case for applications with high dynamic pressure pulsations or vibrations
- For gaseous, dry and aggressive media, also in aggressive ambience
- Process industry: chemical/petro-chemical, pharmaceutical, biotechnology, machine and power generation industries
Special Features:
- Zero adjustment in front
- All stainless steel construction
- Special connection location on request
- Low scale ranges from 0 … 2.5 mbar
Description:
- Design EN 837-3
- Nominal size in mm: 63, 100, 160
- Accuracy class 1.6
- Scale ranges
- NS 63: 0 … 40 mbar to 0 … 600 mbar
- NS 100: 0 … 16 mbar to 0 … 600 mbar
- NS 160: 0 … 2.5 mbar to 0 … 600 mbar
- or all other equivalent vacuum or combined pressure and vacuum ranges
Pressure limitation:
- Steady: Full scale value
- Fluctuating: 0.9 x full scale value
Permissible temperature
- Ambient: -20 … +60 °C
- Medium: +100 °C maximum
Temperature effect: When the temperature of the measuring system deviates from the reference temperature (+20 °C): max. ±0.6 %/10 K of full scale value
Ingress protection per IEC/EN 60529
- IP54 for model 632.50 (without liquid filling)